Z-0854-2007
LADAR6000 Excimer Laser System
Alcon Laboratories / initiated 2007-02-21 / Terminated / root cause: software
Alcon Refractive Horizons, Inc. began a recall of LADAR6000 Excimer Laser System on . The reason FDA records is "Observed 'central islands' in patients following custom myopia laser procedures using the LADAR6000 Excimer Laser. (On May 11, 2007, Alcon notified consignees that the device software programs for CustomCornea Myopia...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0854-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alcon Laboratories
- Product
- LADAR6000 Excimer Laser System
- Product code
- LZS Excimer Laser System
- Reason for recall
- Observed 'central islands' in patients following custom myopia laser procedures using the LADAR6000 Excimer Laser. (On May 11, 2007, Alcon notified consignees that the device software programs for CustomCornea Myopia and CustomCornea Myopia with Astigmatism of the device would be deactivated)
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Consignees were notified of the recall via two letters. On Feb. 21, 2007, consignees were requested to immediately stop performing all custom cornea myopia procedures and to return an acknowledgement card indicating they have received the alert and understand the required actions. On May 11, 2007, Alcon notified consignees that the device software programs for CustomCornea Myopia and CustomCornea Myopia with Astigmatism of the device would be deactivated. After meeting with CDRH's Office of Compliance on August 27, 2007, the recalling firm committed to total product removal of the device. Notification will begin in Sept 2007 and will be made by personal visits to each consignee.
- Quantity in commerce
- 102
- Distribution
- Worldwide Distribution-including USA, Czech Republic, Spain, Australia, Italy, Hong Kong, Vietnam, Singapore, South Korea, and Canada
- Product codes and lots
- All serial/lot numbers of the LADAR6000