Z-1105-2007

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc...

Smith & Nephew, Inc. / initiated 2004-10-12 / Terminated

Smith & Nephew, Inc Endoscopy Division began a recall of Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810 on . The reason FDA records is "Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1105-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.
Product code
DWQ Stripper, Vein, External
Reason for recall
Build-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.
FDA root cause
Other
Action
Firm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.
Quantity in commerce
274 units.
Distribution
Product sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.
Product codes and lots
Pnuematic Footswitch part of TRIVEX System with the following serial numbers: 105, 106, 108, 109, 111 -123, 125 - 182, 184 - 188, 192 - 197, 199 - 203, 205 - 219, 222 - 249, 252 - 254, and 7210386.