Synovis Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement, ( PSD 21-U-V), 21mm...
Synovis Surgical Inovation Div. of began a recall of Synovis Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement, ( PSD 21-U-V), 21mm, US Surgical Staplers. Each package contains the... on . The reason FDA records is "Package Integrity/Sterility Compromised-The foil pouch in the Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement has a potential for minute defects which may compromise the sterility of the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0963-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- SYNOVIS LIFE TECHNOLOGIES, Inc. (A Subsidiary of Baxter International Inc.)
- Product
- Synovis Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement, ( PSD 21-U-V), 21mm, US Surgical Staplers. Each package contains the following components: Cartridge and anvil assemblies, Cartridge cone, PSD Gel, & C-clip
- Product code
- FTM Mesh, Surgical
- Reason for recall
- Package Integrity/Sterility Compromised-The foil pouch in the Peri-Strips Dry with Veritas Collagen Matrix Circular Staple Line Reinforcement has a potential for minute defects which may compromise the sterility of the outside of the tray in which the device is packaged.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Recall notification letters were sent to hospitals via Federal Express on 03/16/2007. This letter describes the product being recalled and problem. It requests consignees to complete and fax back the product return form. EXPANDED Another recall letter dated 04/23/2007 was sent to inform customers of 17 aditional lot numbers being recalled. A third recall letter dated 05/05/2007 was set to hospitals to inform them of one additional Lot/Serial number that was affected. This letter also requests a product return form be completed and faxed to Synovis Surgical Innovations.
- Quantity in commerce
- 85
- Distribution
- Nationwide including states of AZ, CA, CT, FL, IA, IL, IN, KY, MA, MI, MN, MO, NJ, NY, OH, OK, PA, TN, TX, VA, and WI
- Product codes and lots
- 5713057-221865, 5713421-227240, 5727945-538783, 5728518-547271. EXPANDED SERIAL/LOT#'s: 5728518-560761 and 5729324-562206
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
