TDx/TDxFLx Methotrexate II Reagent-List Number 07A12-60 - a reagent system for In Vitro diagnostic use. This...
Abbott Diagnostic International, Ltd. / initiated 2007-03-20 / Terminated
Abbott Diagnostic International, Ltd. began a recall of TDx/TDxFLx Methotrexate II Reagent-List Number 07A12-60 - a reagent system for In Vitro diagnostic use. This reagent system is composed of one bottle of Wash... on . The reason FDA records is "Incorrect configuration: kits contain reagent bottles in the order S-W-T-P instead of the correct order, W-S-T-P". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1056-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostic International, Ltd.
- Product
- TDx/TDxFLx Methotrexate II Reagent-List Number 07A12-60 - a reagent system for In Vitro diagnostic use. This reagent system is composed of one bottle of Wash Solution, one bottle of Methotrexate Antibody, one bottle of Methotrexate Fluorescein Tracer and one bottle of Pretreatment Solution.
- Product code
- LAO Enzyme Immunoassay, Methotrexate
- Reason for recall
- Incorrect configuration: kits contain reagent bottles in the order S-W-T-P instead of the correct order, W-S-T-P.
- FDA root cause
- Other
- Action
- The Product Correction Letter dated 3/20/2007, was sent to all customers who received the affected lots. The Customers were instructed to: Determine if they had the lot listed in the Product Correction letter. Inspect each kit of the affected lot and discard any kits incorrectly configured. If they had forwarded the lots listed in the Product Recall letter to other laboratories, they were instructed to inform them of this Product Recall and provide a copy of this letter to them and retain Product Correction letter for their laboratory records.
- Quantity in commerce
- 1,026
- Distribution
- World wide Distribution : USA including states of AR, CA, CO, CT, DE, FL,GA, ID, IL, IA, KS, KY, LA, MA, ME, MD, MI, MN, MS, MO, NJ, NY, NC,NV, OH, SC, PA, RI, TN, TX, UT, WA, WI, WV, Hawaii, Washington DC, and Puerto Rico. Countries of : Canada, Mexico, Venezuela, Columbia, Chile, Brazil, Germany, Japan, Hong Kong, Australia and New Zealand
- Product codes and lots
- Lot Number: 45050Q100