Z-0750-2007
Mobilett XP, mobile X-ray system, Model number 1818447
Siemens Medical Solutions USA, Inc. / initiated 2007-03-21 / Terminated
Siemens Medical Solutions USA, Inc began a recall of Mobilett XP, mobile X-ray system, Model number 1818447 on . The reason FDA records is "Mounting bolts for tank fork assembly and collimator flange may become loose". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0750-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mobilett XP, mobile X-ray system, Model number 1818447
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Mounting bolts for tank fork assembly and collimator flange may become loose.
- FDA root cause
- Other
- Action
- Siemens Service Personnel are visiting each consignee, beginning March 21, 2007, to replace the required mounting bolts and tighten them according to required torque specifications.
- Quantity in commerce
- 36 units
- Distribution
- The products were shipped to medical facilities nationwide.
- Product codes and lots
- Serial numbers 2033, 1034, 1010, 1020, 1041, 1053, 2044, 2045, 1008, 1043, 1025, 1026, 1027, 1028, 1029, 1030, 1031, 1032, 1033, 1040, 2007, 1052, 2001, 1017, 2002, 2040, 2021, 2022, 1011, 1016, 1018, 1021, 1039, 1054, 2005, and 2017.