Z-1005-2007

Terumo Perfusion System 8000 Roller Pump, 1040V; Model 8000; Catalog No. 16411. (Not distributed within the...

Terumo Cardiovascular Systems Corporation / initiated 2004-07-01 / Terminated

Terumo Cardiovascular Systems Corp began a recall of Terumo Perfusion System 8000 Roller Pump, 1040V; Model 8000; Catalog No. 16411. (Not distributed within the United States) on . The reason FDA records is "Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1005-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Perfusion System 8000 Roller Pump, 1040V; Model 8000; Catalog No. 16411. (Not distributed within the United States).
Product code
DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Reason for recall
Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
FDA root cause
Other
Action
As the complaints were received (between 07/04 and 01/06), the firm visited the customer filing the complaint and repaired the pumps demonstrating the problem. The firm will correct other units as complaints are received. The firm modified its recall strategy and issued a Device Correction Letter dated 6/22/07 notifying all users of the problem
Quantity in commerce
728 pumps of all models
Distribution
Worldwide: USA and Bangladesh, Belgium, Bolivia, Canada, Chile, Colombia, Costa Rica, Egypt, Equador, Hong Kong, India, Indonesia, Japan, Latin America, Malaysia, Nicaragua, Philippines, Saudi Arabia, Taipe, Taiwan, United Arab Emirates, Uruguay, and Viet Nam.
Product codes and lots
Serial numbers 10031, 10033, 10137, 10138, 10614, 10663, 10730, 11131, 11281, 11283, 11361, 11548, 11630, 11751, 12110, 12604, 12686, 12938, 13129, 13719, 13826, 13830, 14712, 15090, 15277, 15278, 15279, 15280, 15284, 15285, 15304, 16003, 16087, 16190, 17099, 17345, 17432 and 17437 through 18122 may be affected by the problem. None have been corrected.