Z-0772-2007
Lumex Walkabout Lite Four-Wheel Rollator, Manufactured for: Graham-Field Health Products, Atlanta, Georgia...
GF Health Products, Inc. / initiated 2007-04-02 / Terminated
Graham Field Atlanta began a recall of Lumex Walkabout Lite Four-Wheel Rollator, Manufactured for: Graham-Field Health Products, Atlanta, Georgia 30360 on . The reason FDA records is "The caster bolt stem/screw may loosen over time causing the wheel to come off". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0772-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GF Health Products, Inc.
- Product
- Lumex Walkabout Lite Four-Wheel Rollator, Manufactured for: Graham-Field Health Products, Atlanta, Georgia 30360.
- Product code
- ITJ Walker, Mechanical
- Reason for recall
- The caster bolt stem/screw may loosen over time causing the wheel to come off.
- FDA root cause
- Other
- Action
- Graham-Field notified all of its Durable Medical Equipment (DME) dealers involved in the distribution of these products by certified US Mail on April 3, 2007. They were made aware of the need to adjust the product to lessen the risk of injury and were provided a re-work kit. The distribution of DME dealers includes the announcement of the field correction to the DME dealer, the letter to send to the end-users, rework supplies, rework instructions and proof of compliance cards. Graham-Field instructed the DME dealers to adjust any impacted products on site, as well as, any delivered by end-users, customers or other impacted third parties to the DME dealer for correction. DME dealers will be compensated for performing the corrective action. The installation of the rework kit involves the installation of lock washers and Locitite and is described in more detail in the attached rework instructions. Graham-Field does not have direct access to end user names, however it plans to reach end-users by financially supporting a certified mailing from DME dealers to their individual customers notifying them of this field correction. A follow-up 60 day certified mailing will be performed for non-responders.
- Quantity in commerce
- 13,300 units
- Distribution
- U.S. Nationwide, Costa Rica, Ecuador, El Salvador, Guatemala, and Santo Domingo
- Product codes and lots
- Model numbers: RJ4300B, RJ4300D, RF4300G, RJ4300K, RJ4300R, and Imperial Model RF4400K. (Note: Models distributed after January 2007, serial numbers 120791 through 145254, are not subject to this correction)