These models are not available in the US. CONTACT RENEWAL 4 AVT HE (model number M170, M175, M177, M179)...
Boston Scientific CRM Corp began a recall of These models are not available in the US. CONTACT RENEWAL 4 AVT HE (model number M170, M175, M177, M179). Cardiac Resynchronization Therapy Defibrillator... on . The reason FDA records is "Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may...". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-1051-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- These models are not available in the US. CONTACT RENEWAL 4 AVT HE (model number M170, M175, M177, M179). Cardiac Resynchronization Therapy Defibrillator. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Product code
- NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Reason for recall
- Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.
- Root cause tag
- supplier
- FDA root cause
- Other
- Action
- An Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.
- Quantity in commerce
- 539
- Distribution
- Worldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Polynesia, Portugal, Puerto Rico, Qatar, Saint Lucia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. Virgin Islands, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, and West Indies.
- Product codes and lots
- Model M170, serial numbers: 100136, 100138, 100142, 100144, 100145, 200003, 200004, 200006, 200007, 200009, 200010, 200011, 200013, 200017, 200018, 200019, 200021, 200022, 200023, 200025, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200041, 200042, 200043, 200046, 200047, 200048, 200051, 200052, 200053, 200054, 200056, 200057, 200058, 200059, 200060, 200073, 200074, 200077, 200079, 200082, 200084, 200087, 200088, 200089, 200090, 200093, 200094, 200095, 200098, 200101, 200112. Model M175, serial numbers: 100202, 100207, 100209, 100210, 100211, 100212, 100213, 100214, 100215, 100216, 100219, 100221, 100222, 100224, 100251, 100277, 100282, 100283, 100284, 100298, 100299, 100336, 100341, 100344, 100345, 100349, 100350, 100353, 100354, 100356, 100358, 100361, 100362, 100369, 100370, 100372, 100384, 100401, 100404, 100407, 100409, 100410, 100411, 100418, 100421, 100433, 200002, 200003, 200004, 200005, 200007, 200009, 200010, 200011, 200012, 200014, 200015, 200016, 200017, 200019, 200022, 200025, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200036, 200038, 200039, 200040, 200042, 200044, 200047, 200048, 200051, 200052, 200053, 200054, 200056, 200057, 200058, 200060, 200062, 200067, 200071, 200074, 200075, 200079, 200080, 200081, 200083, 200084, 200085, 200086, 200088, 200089, 200091, 200092, 200093, 200094, 200095, 200108, 200112, 200113, 200115, 200124, 200143, 200153, 200156, 200157, 200158, 200162, 200163, 200166. Model M177, serial numbers: 100154, 100157, 100158, 100168, 100169, 100173, 100174, 100180, 100181, 100182, 100183, 100190, 100191, 100196, 100197, 100221, 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200011, 200012, 200013, 200014, 200015, 200016, 200017, 200018, 200020, 200021, 200022, 200023, 200024, 200025, 200026, 200027, 200028, 200029, 200030, 200032, 200033, 200034, 200035, 200036, 200037, 200038, 200039, 200040, 200041, 200042, 200044, 200045, 200046, 200047, 200048, 200049, 200052, 200053, 200054, 200055, 200056, 200057, 200058, 200059, 200060, 200062, 200066, 200067, 200068, 200069, 200072, 200073, 200074, 200076, 200078, 200080, 200081, 200083, 200084, 200085, 200086, 200087, 200089, 200090, 200091, 200092, 200093, 200095, 200096, 200097, 200099, 200102, 200104, 200105, 200106, 200107, 200108, 200109, 200110, 200111, 200112, 200113, 200114, 200115, 200116, 200117, 200118, 200119, 200120, 200121, 200122, 200123, 200124, 200125, 200126, 200128, 200129, 200130, 200131, 200134, 200135, 200136, 200140, 200141, 200142, 200143, 200145, 200146, 200147, 200148, 200149, 200150, 200152, 200153, 200154, 200155, 200156, 200158, 200160, 200161, 200164, 200165, 200166, 200167, 200168, 200169, 200170, 200171, 200172, 200175, 200177, 200178, 200179, 200180, 200181, 200182. Model M179, serial numbers: 100216, 100217, 100219, 100223, 100227, 100228, 100229, 100230, 100231, 100232, 100233, 100235, 100237, 100238, 100239, 100240, 100243, 100244, 100247, 100250, 100253, 100256, 100257, 100263, 100266, 100270, 100271, 100272, 100274, 100275, 100276, 100278, 100279, 100280, 100281, 100282, 100283, 100285, 100287, 100290, 100291, 100293, 100294, 100295, 100297, 100298, 100300, 100302, 100303, 100304, 100305, 100315, 100316, 100317, 100319, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100338, 100342, 100343, 100349, 100353, 100354, 100391, 100430, 100435, 100436, 100437, 100439, 100440, 100441, 100442, 100448, 100449, 100450, 100452, 100467, 100469, 100470, 100473, 100475, 100479, 100480, 100482, 100488, 100489, 100500, 100502, 100503, 100515, 100516, 100517, 100520, 100522, 100533, 100543, 100545, 100547, 100549, 100552, 100555, 100556, 100557, 100558, 100563, 100566, 100567, 100568, 100577, 100579, 100581, 100584, 100587, 100589, 100590, 100591, 100614, 200002, 200003, 200004, 200005, 200006, 200007, 200008, 200009, 200010, 200013, 200014, 200015, 200016, 200017, 200018, 200019, 200020, 200021, 200022, 200023, 200024, 200026, 200027, 200028, 200029, 200030, 200031, 200032, 200035, 200036, 200037, 200039, 200040, 200042, 200043, 200044, 200046, 200047, 200049, 200051, 200052, 200053, 200054, 200055, 200056, 200058, 200059, 200060, 200061, 200062, 200063, 200064, 200065, 200066, 200068, 200070, 200071, 200072, 200075, 200077, 200078, 200079, 200080, 200081, 200083, 200084, 200086, 200088, 200089, 200090, 200092, 200093, 200094, 200096, 200097, 200100, 200101.