Z-0792-2007

ARCHITECT Folate Reagent, for in vitro diagnostics, List Numbers 07K60-20 (4 x 100 test kits), 07K60-25 (100...

Abbott Medical / initiated 2007-03-28 / Terminated / root cause: manufacturing process

Abbott Laboratories, Inc began a recall of ARCHITECT Folate Reagent, for in vitro diagnostics, List Numbers 07K60-20 (4 x 100 test kits), 07K60-25 (100 test kits)and 07K60-30 (4 x 500 test kits)... on . The reason FDA records is "If the ARCHITECT i2000/i2000sr System becomes contaminated with microbes generating folate-like by-products, the ARCHITECT Folate assay may experience calibration failures and/or shifts in results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0792-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT Folate Reagent, for in vitro diagnostics, List Numbers 07K60-20 (4 x 100 test kits), 07K60-25 (100 test kits)and 07K60-30 (4 x 500 test kits); manufactured by Abbott Ireland Diagnostic Division, Lisnamuck, Longford, Ireland.
Product code
CGN Acid, Folic, Radioimmunoassay
Reason for recall
If the ARCHITECT i2000/i2000sr System becomes contaminated with microbes generating folate-like by-products, the ARCHITECT Folate assay may experience calibration failures and/or shifts in results.
Root cause tag
manufacturing process
FDA root cause
Other
Action
Notification Letters were sent to all customers who received the ARCHITECT Folate Reagent on 03/28/07. Customers were provided with instructions on how to conduct a Wash Buffer Unload procedure and were requested to perform this action daily. Customers were also asked to contact their local Abbott Customer Support Rep to schedule an Internal Decontamination procedure on the ARCHITECT System. Firm recalling to the user (laboratory) level.
Quantity in commerce
3866 units.
Distribution
Folate Reagent distributed to the following countries: Australia, Austria, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Finland, France, Greece, Hungary, Ireland, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Oman, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia and Montenegro, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, UAE, United Kingdom. No US distribution.
Product codes and lots
Product Control/Lot # (For List #07K60-20): 42908JN00, 42964JN00 and 42905JN00; (For List #07K60-25): 42909JN00, 42965JN00 and 42906JN00; (For List #07K60-30): 42910JN00, 42966JN00, 42907JN00, and 42931JN00.