Z-0920-2007

Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog...

Roche Diagnostics GmbH / initiated 2007-04-30 / Terminated

Roche Diagnostics Corp. began a recall of Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190 on . The reason FDA records is "Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0920-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.
Product code
LEG Enzyme Immunoassay, Valproic Acid
Reason for recall
Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
FDA root cause
Other
Action
Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
Quantity in commerce
992
Distribution
Nationwide.
Product codes and lots
Lots 144107, exp. 9/30/2007; 144166, exp. 1/31/2008, and 144208, exp. 6/30/2008