Z-0920-2007
Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog...
Roche Diagnostics GmbH / initiated 2007-04-30 / Terminated
Roche Diagnostics Corp. began a recall of Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190 on . The reason FDA records is "Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0920-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche VALP2, Online TDM Valproic acid , for use with the Roche/Hitachi Cobas c 501 analyzer; Catalog No.04491041190.
- Product code
- LEG Enzyme Immunoassay, Valproic Acid
- Reason for recall
- Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
- FDA root cause
- Other
- Action
- Consignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
- Quantity in commerce
- 992
- Distribution
- Nationwide.
- Product codes and lots
- Lots 144107, exp. 9/30/2007; 144166, exp. 1/31/2008, and 144208, exp. 6/30/2008