Z-0839-2007
Silver Speed Hydrophilic Guidewire, for placement of vascular catheters. Model # 103-0601-200
Medtronic, Inc. / initiated 2007-04-05 / Terminated
EV3 Neurovascular began a recall of Silver Speed Hydrophilic Guidewire, for placement of vascular catheters. Model # 103-0601-200 on . The reason FDA records is "Two customer complaints were received from physicians in Japan indicating the hydrophilic coating was observed flaking from the guidewire while immersed in hydrating media following use during aneurysm embolization...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0839-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Silver Speed Hydrophilic Guidewire, for placement of vascular catheters. Model # 103-0601-200.
- Product code
- MKI
- Reason for recall
- Two customer complaints were received from physicians in Japan indicating the hydrophilic coating was observed flaking from the guidewire while immersed in hydrating media following use during aneurysm embolization procedures.
- FDA root cause
- Other
- Action
- A recall plan was developed for 100% recovery of guidewires within affected lot numbers. The plan includes notification to U.S., European, and Rest-of-World customers and distributors. A recall notification letter was prepared 04/05/2007 which summarizes the coating problem and potential risks, identifies the affected lot numbers and quantities shipped to the customer, and provides instructions for return of unused guidewires.
- Quantity in commerce
- 2120
- Distribution
- Worldwide, including USA, Italy, United Kingdom, Switzerland, Germany, Belgium, South Africa, Canada, Slovenia, China, Austria, France, Spain, Jordan, Mexico, Portugal, Colombia, Greece, Brazil, Venezula, Ireland, Australia, Malaysia, Saudi Arabia, South Korea, Lithuania, Netherlands, Hungary, and United Arab Emirates
- Product codes and lots
- US: Lot # : 2295725; 2382836; 2355074; 2390635; 2348116; 2355074; 2374619; 2216295; 2257803;2390635; 2342166; ;2390635;2348116; 2310664; 2382836; 2257803; 2216295; 2310664; 2390635; 2247199; 2247200; 2221847; 2247200; 2295725; 2310664; 2221845; 2310664; 2216296; 2257803; 2342166; European Union & Rest-Of-World Lot #: 2374619 ; 2209708;2216295; 2247200; 2310664; 2348116; 2221845 ; 2355074; 2247200; 2310664; 2216295; 2166877; 2247199; 2209708; 2367901; 2370509; 2374619; 2216295; 2163496; 2174470; 2295725; 2370509; 2376907; 2209708; 2295725; 2310664; 2221845; 2247200; 2216291; 2390635; 2360938; 2370509; 2376907; 2310664; 2355074; 2163496; 2276405; 2370508; 2221847; 2257803; 2390635; 2257803; 2370509; 2200727;