Z-0816-2007
Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX...
Palodex Group Oy / initiated 2006-04-28 / Terminated
Instrumentarium Dental, Inc. began a recall of Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland on . The reason FDA records is "There may be a material defect in the up/down movement axle of the unit. This may cause the upper part of the unit to suddenly come down". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0816-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Palodex Group Oy
- Product
- Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland.
- Product code
- EDH
- Reason for recall
- There may be a material defect in the up/down movement axle of the unit. This may cause the upper part of the unit to suddenly come down.
- FDA root cause
- Other
- Action
- A recall letter, dated 04/28/2006, was sent to affected customers . The letter describes the issue, and requests that no patient operation is permitted until the unit is checked by authorized service technician. If the service technician noted wear of the screw, the assembly was replaced.
- Quantity in commerce
- 9
- Distribution
- Worldwide, including USA, Canada, Bermuda, and Colombia.
- Product codes and lots
- Serial numbers X51146, X51147, X51148, X51154, X51155, X51156, X51170, X51171, X61188