SOPRO Camera Control Units, PAL, SV420P, REF (Catalog) #72200506 v1.1; Made in France; Manufactured for Smith...
Smith & Nephew, Inc Endoscopy Division began a recall of SOPRO Camera Control Units, PAL, SV420P, REF (Catalog) #72200506 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810 on . The reason FDA records is "Surgical camera control units were not designed and manufactured to pass the IEC 60101 testing for electrical fast transients (EFT)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0782-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- SOPRO Camera Control Units, PAL, SV420P, REF (Catalog) #72200506 v1.1; Made in France; Manufactured for Smith & Nephew, Inc., Endoscopy, Andover, MA 01810.
- Product code
- FWF Camera, Television, Endoscopic, Without Audio
- Reason for recall
- Surgical camera control units were not designed and manufactured to pass the IEC 60101 testing for electrical fast transients (EFT)
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recalling firm sent Recall notification letters to customers (Jan 2007) and distribution centers requesting device return.
- Quantity in commerce
- 25 units.
- Distribution
- Foreign Distribution Only-No U. S. A. distribution International distribution to: Argentina, Austria, Australia, Belgium, Brazil, China, Columbia, Denmark, Dubai, France, Germany, Greece, Italy, India, Japan, Korea, Malaysia, Netherlands, Norway, Portugal, South Africa, Spain, Sweden, Turkey and United Kingdom.
- Product codes and lots
- Serial #s: 420-8283, 420-8275, 420-8285, 420-9006, 420-8271, 420-8277, 420-8289, 420-9015, 420-9000, 420-9001, 420-9002, 420-9004, 420-9005, 420-9007, 420-9008, 420-9009, 420-9010, 420-9012, 420-9014, 420-9016, 420-9017, 420-9018, 420-9019, 420-8282 and 420-9013.