Z-0824-2007
BinaxNOW RSV 10 Test Kit Part Number : 430-100, contains NP Swabs with '8995' labeled on the swab wrapper
Abbott Diagnostics Scarborough, Inc. / initiated 2007-03-28 / Terminated
Binax, Inc. dba IMPD began a recall of BinaxNOW RSV 10 Test Kit Part Number : 430-100, contains NP Swabs with '8995' labeled on the swab wrapper on . The reason FDA records is "Tip of the foam swab could detach from the shaft causing injury to the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0824-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostics Scarborough, Inc.
- Product
- BinaxNOW RSV 10 Test Kit Part Number : 430-100, contains NP Swabs with '8995' labeled on the swab wrapper
- Product code
- GQG Antigen, Cf (Including Cf Controls), Respiratory Syncytial Virus
- Reason for recall
- Tip of the foam swab could detach from the shaft causing injury to the patient.
- FDA root cause
- Other
- Action
- Inverness Medical issued recall notification by letter on 3/28/07. Customers are instructed to discard the NP swabs provided in the kits. The firm will replace the NP swabs. In the meantime customers are to use the tests with nasal wash/aspirate samples (influenza) or nasal wash samples (RSV). Response letters are to be signed by customers and returned to Inverness Medical.
- Quantity in commerce
- 2851 kits
- Distribution
- Nationwide and Japan.
- Product codes and lots
- Kit Lot Numbers: 023437, 024056, 024057, 024436, 024516, 024518, 024694, 024695, 024882, 025090, 025089 NP Swabs Lot Number: 8995