Z-0811-2007

Certain MicroMiniplant Straight Healing Abutment. Dental implant. Model numbers: ISMHA32, ISMHA33, ISMHA34...

ZimVie US Corp LLC / initiated 2007-04-25 / Terminated

Biomet 3i began a recall of Certain MicroMiniplant Straight Healing Abutment. Dental implant. Model numbers: ISMHA32, ISMHA33, ISMHA34 and ISMHA36 on . The reason FDA records is "The Straight Healing Abutment did not fully seat to the restorative platform of the implant, leaving a gap between the abutment and the implant". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0811-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Certain MicroMiniplant Straight Healing Abutment. Dental implant. Model numbers: ISMHA32, ISMHA33, ISMHA34 and ISMHA36.
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
The Straight Healing Abutment did not fully seat to the restorative platform of the implant, leaving a gap between the abutment and the implant.
FDA root cause
Other
Action
Each consignee was notified on 4/25/2007 by telephone using a prepared script. A follow letter was sent by fax. Each consignee was instructed to check their respective inventory for the affected product and to return a response card and any affected product to the firm.
Quantity in commerce
362 units
Distribution
Distributed worldwide, including USA, Spain and the Netherlands.
Product codes and lots
lot Numbers: 593025-5, 593027-5, 598408-5, 593028-5, 599866-5 and 593029-5.