Z-0813-2007

BD 5ml Syringe Luer-Lok Tip Bulk Sterile Convenience Pak Latex Free

Becton Dickinson & Co / initiated 2007-04-09 / Terminated / root cause: packaging

Becton Dickinson & Company began a recall of BD 5ml Syringe Luer-Lok Tip Bulk Sterile Convenience Pak Latex Free on . The reason FDA records is "Small cracks in plastic trays of one lot of BD 5mL Syringe Luer Lok Tip, Bulk Sterile Convenience Paks that may adversely affect tray integrity and steriity". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0813-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Becton Dickinson & Co
Product
BD 5ml Syringe Luer-Lok Tip Bulk Sterile Convenience Pak Latex Free
Product code
FMF Syringe, Piston
Reason for recall
Small cracks in plastic trays of one lot of BD 5mL Syringe Luer Lok Tip, Bulk Sterile Convenience Paks that may adversely affect tray integrity and steriity.
Root cause tag
packaging
FDA root cause
Other
Action
Recall communications were sent on April 2, 2007 and May 3, 2007 by UPS second day air to all direct consignees and potential customers.
Quantity in commerce
174,300 units
Distribution
Nationwide.
Product codes and lots
Lot number 7002604 Ref 309703