Z-0825-2007
BinaxNOW Nasopharyngeal (NP) Accessory Pak, Part Number 400-065, contain NP Swabs with '8995' labeled on the...
Abbott Diagnostics Scarborough, Inc. / initiated 2007-03-28 / Terminated
Binax, Inc. dba IMPD began a recall of BinaxNOW Nasopharyngeal (NP) Accessory Pak, Part Number 400-065, contain NP Swabs with '8995' labeled on the swab wrapper on . The reason FDA records is "Tip of the foam swab could detach from the shaft causing injury to the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0825-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostics Scarborough, Inc.
- Product
- BinaxNOW Nasopharyngeal (NP) Accessory Pak, Part Number 400-065, contain NP Swabs with '8995' labeled on the swab wrapper
- Product code
- GNX Antigens, Cf (Including Cf Control), Influenza Virus A, B, C
- Reason for recall
- Tip of the foam swab could detach from the shaft causing injury to the patient.
- FDA root cause
- Other
- Action
- Inverness Medical issued recall notification by letter on 3/28/07. Customers are instructed to discard the NP swabs provided in the kits. The firm will replace the NP swabs. In the meantime customers are to use the tests with nasal wash/aspirate samples (influenza) or nasal wash samples (RSV). Response letters are to be signed by customers and returned to Inverness Medical.
- Quantity in commerce
- 1101 kits
- Distribution
- Nationwide and Japan.
- Product codes and lots
- Kit lot Numbers: 023357, 023566, 024629, 024631,
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
