Z-0826-2007
BinaxNOW A&B Kit 10 Test Kit (Japan) Part No. 416-110J contains NP Swabs with '8995' labeled on the swab...
Abbott Diagnostics Scarborough, Inc. / initiated 2007-03-28 / Terminated
Binax, Inc. dba IMPD began a recall of BinaxNOW A&B Kit 10 Test Kit (Japan) Part No. 416-110J contains NP Swabs with '8995' labeled on the swab wrapper on . The reason FDA records is "Tip of the foam swab could detach from the shaft causing injury to the patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0826-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostics Scarborough, Inc.
- Product
- BinaxNOW A&B Kit 10 Test Kit (Japan) Part No. 416-110J contains NP Swabs with '8995' labeled on the swab wrapper
- Product code
- GNX Antigens, Cf (Including Cf Control), Influenza Virus A, B, C
- Reason for recall
- Tip of the foam swab could detach from the shaft causing injury to the patient.
- FDA root cause
- Other
- Action
- Inverness Medical issued recall notification by letter on 3/28/07. Customers are instructed to discard the NP swabs provided in the kits. The firm will replace the NP swabs. In the meantime customers are to use the tests with nasal wash/aspirate samples (influenza) or nasal wash samples (RSV). Response letters are to be signed by customers and returned to Inverness Medical.
- Quantity in commerce
- 3814 kits
- Distribution
- Nationwide and Japan.
- Product codes and lots
- Lot number: 24162