Z-1039-2007

AxSYM HAVAB 2.0 Reagent Pack; list 08D21; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott...

Abbott Medical / initiated 2007-04-23 / Terminated

Abbott Laboratories began a recall of AxSYM HAVAB 2.0 Reagent Pack; list 08D21; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064 on . The reason FDA records is "Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1039-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
AxSYM HAVAB 2.0 Reagent Pack; list 08D21; 100 test pack; in-vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064
Product code
LOL Hepatitis A Test (Antibody And Igm Antibody)
Reason for recall
Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
FDA root cause
Other
Action
Abbott initiated a field correction recall by letter dated 4/23/07 to all customers receiving the affected reagents. The accounts were informed of the failure of the AxSYM to open (actuate) some reagent packs with the affected lots. The accounts were requested to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading the reagent pack on the instrument; and discard any reagent packs that do not open properly. If the user observes a probe crash, prior to testing samples again, the user must perform a probe crash recovery following the procedure in the Operations Manual. If necessary, they need to replace the probe. If a probe crash recovery is not performed, subsequent results may be impacted. The customers were requested to retain a copy of the letter for their records. Customers requiring additional information were directed to contact Abbott Customer Service at 1-877-422-2688.
Quantity in commerce
1,111 packs
Distribution
Worldwide, including USA, Puerto Rico, Argentina, Australia, Brazil, Canada, Cayman Islands, Chile, Colombia, Ecuador, Germany, Hong Kong, Mexico, New Zealand, Singapore, South Korea, Taiwan, Thailand and Venezuela.
Product codes and lots
list 08D21-20; lots 48132M200, 48139M200, 51069M100