Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0172-2008

Cyberonics VNS Therapy Bipolar Lead, Model #302; Sterile, Single use only; Distributed by Cyberonics, Inc...

LivaNova USA, Inc. / initiated 2007-08-02 / Terminated / root cause: manufacturing process

Cyberonics, Inc began a recall of Cyberonics VNS Therapy Bipolar Lead, Model #302; Sterile, Single use only; Distributed by Cyberonics, Inc., Houston, TX 77058 on . The reason FDA records is "Dissolution/Fractures to the leads of the VNS Therapy System". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0172-2008
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Cyberonics VNS Therapy Bipolar Lead, Model #302; Sterile, Single use only; Distributed by Cyberonics, Inc., Houston, TX 77058.
Product code
LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
Reason for recall
Dissolution/Fractures to the leads of the VNS Therapy System
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Cyberonics issued a correction notice (Safety Alert) to physicians in June of 2007, informing them of the present labeling and including updated labeling to provide instructions on how to detect lead fractures and how to minimize risk to patients.
Quantity in commerce
34,719 units.
Distribution
Worldwide Distribution - USA and countries of Argentina, Australia, Austria, Belgium, Canada, Costa Rica, Cyprus, Czech Rep, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Jordan, Korea, Malta, Mexico, Netherlands, Northern Ireland, Norway, Poland, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom, and Venezuela.
Product codes and lots
All manufactured and distributed leads.