AxSYM Total B-hCG Reagent List 7A59-21 is composed of one bottle of Anti B-hCG Coated Microparticles, one...
Abbott Diagnostic International, Ltd. began a recall of AxSYM Total B-hCG Reagent List 7A59-21 is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti- B-hCG Alkaline Phosphatase Conjugate... on . The reason FDA records is "An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1067-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostic International, Ltd.
- Product
- AxSYM Total B-hCG Reagent List 7A59-21 is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti- B-hCG Alkaline Phosphatase Conjugate and one bottle of Specimen Diluent. REF 7A59-21. 100 Tests. For In Vitro Diagnostic Use.
- Product code
- JHI Visual, Pregnancy Hcg, Prescription Use
- Reason for recall
- An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.
- FDA root cause
- Other
- Action
- The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots.
- Quantity in commerce
- 11530 kits
- Distribution
- Worldwide, including USA, PR, USVI, CANADA, MEXICO, GUATEMALA, COSTA RICA, COLOMBIA, ECUADOR, PERU, CHILE, URUGUAY, ARGENTINA, BRAZIL, DOMINICAN REPUBLIC, EL SALVADOR, GERMANY, JAPAN, SINGAPORE, HONG KONG, TAIWAN, THAILAND, SOUTH KOREA, AUSTRALIA, NEW ZEALAND, HONDURAS, TRINIDAD & TOBAGO, GRENADA, ST. VINCENT, ST. LUCIA, ANTIGUA, BARBADOS, CURACAO, BRITISH VIRGIN ISLANDS, BAHAMAS, PANAMA, and CAYMAN ISLANDS.
- Product codes and lots
- 46329Q101, 47118Q101, 48063Q101, 48065Q101, 49341Q101, 49559Q101