MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001
Roche Diagnostics Corp. began a recall of MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001 on . The reason FDA records is "False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0956-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.
- Product code
- NPQ Test, Factor V Leiden Mutations, Genomic Dna Pcr
- Reason for recall
- False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.
- Quantity in commerce
- 4,201
- Distribution
- Nationwide.
- Product codes and lots
- Lots 11962700, exp. 8/31/07; 12030400, exp. 6/30/07; 12071100, exp. 9/30/07; 12125800, exp. 10/31/07; 12214300, exp. 12/31/07; 12360200, exp. 3/31/08; 12426200, exp. 4/3008; 12543600, exp. 5/31/08; 12672500, exp. 5/31/08; 12725300, exp. 6/30/08; 12799800, exp. 7/31/08; 12917900, exp. 8/31/08; 13024000, exp. 9/30/08; 13081700, exp. 9/30/08; 13197600, exp. 12/31/08 and 13266600, exp. 2/28/09.