Z-1072-2007

STA Liatest Free Protein S For in vitro diagnostic use. 2-8 degrees C Distributed in the USA by: Diagnostica...

Diagnostica Stago, Inc. / initiated 2007-05-08 / Terminated

Diagnostica Stago, Inc. began a recall of STA Liatest Free Protein S For in vitro diagnostic use. 2-8 degrees C Distributed in the USA by: Diagnostica Stago, Inc. 5 Century Drive Parsippany, NJ Phone... on . The reason FDA records is "Specific lots of STA Liatest Free Protein S may result in low results in the assigned value ranges and sometimes outside the lower limit". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1072-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Diagnostica Stago, Inc.
Product
STA Liatest Free Protein S For in vitro diagnostic use. 2-8 degrees C Distributed in the USA by: Diagnostica Stago, Inc. 5 Century Drive Parsippany, NJ Phone 800-abc coag 1 - 6 x 5 ml Buffer 2 - 6 x 6 ml Latex
Product code
GGP Test, Qualitative And Quantitative Factor Deficiency
Reason for recall
Specific lots of STA Liatest Free Protein S may result in low results in the assigned value ranges and sometimes outside the lower limit.
FDA root cause
Other
Action
Customers were sent a Product Recall Notification on May 8, 2007 by first class mail labeled " Recall notice, Open Immediately"
Quantity in commerce
365 units
Distribution
Medical facilities nationwide
Product codes and lots
Catalog number: 0516 Lot Numbers: 061312, Exp 5/31/07;  061991, Exp 8/31/07;  062633, Exp 10/31/07;  063281, Exp 12/31/07