Z-0988-2007
MediSpectra Luma Cervical Imaging System Model Number: 3-02600
Medispectra Inc. / initiated 2007-05-15 / Terminated
Medispectra Inc. began a recall of MediSpectra Luma Cervical Imaging System Model Number: 3-02600 on . The reason FDA records is "Power supply may fail and render the unit inoperable". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0988-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medispectra Inc.
- Product
- MediSpectra Luma Cervical Imaging System Model Number: 3-02600
- Product code
- MWM Sensor, Electro-Optical (For Cervical Cancer)
- Reason for recall
- Power supply may fail and render the unit inoperable.
- FDA root cause
- Other
- Action
- MediSpectra notified users by telephone on 5/15/07 and follow-up letter. Users are requested to select a date for field servce to replace the power supply.
- Quantity in commerce
- 8 units
- Distribution
- AZ, CA, FL, & NY.
- Product codes and lots
- Serial Numbers: L0054, L0057, L0059, L0062