Z-0988-2007

MediSpectra Luma Cervical Imaging System Model Number: 3-02600

Medispectra Inc. / initiated 2007-05-15 / Terminated

Medispectra Inc. began a recall of MediSpectra Luma Cervical Imaging System Model Number: 3-02600 on . The reason FDA records is "Power supply may fail and render the unit inoperable". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0988-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medispectra Inc.
Product
MediSpectra Luma Cervical Imaging System Model Number: 3-02600
Product code
MWM Sensor, Electro-Optical (For Cervical Cancer)
Reason for recall
Power supply may fail and render the unit inoperable.
FDA root cause
Other
Action
MediSpectra notified users by telephone on 5/15/07 and follow-up letter. Users are requested to select a date for field servce to replace the power supply.
Quantity in commerce
8 units
Distribution
AZ, CA, FL, & NY.
Product codes and lots
Serial Numbers: L0054, L0057, L0059, L0062