Z-1106-2007

Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests

Roche Molecular Systems Inc / initiated 2007-05-21 / Terminated / root cause: labeling and UDI

Roche Molecular Systems, Inc. began a recall of Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests on . The reason FDA records is "For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1106-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Molecular Systems Inc
Product
Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests
Product code
JJF Analyzer, Chemistry, Micro, For Clinical Use
Reason for recall
For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by AMPLILINK Software (versions 1.1, 1.3, 1.4 and 2.41) and the stability information provided in the test kit package inserts/method manual.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Urgent Medical Device Correction letters were mailed on 5/21/07 by first class mail. Letters describe the discrepancy between software reagent expiration dates and package insert expiration dates.
Quantity in commerce
5,806 total
Distribution
Nationwide to hospitals and laboratories.
Product codes and lots
Catalog number 20757497122, working probe suspension /IC PS1-IC4; Catalog Number 20757535122, Working substrate SB3-SB