Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests
Roche Molecular Systems, Inc. began a recall of Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests on . The reason FDA records is "For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1106-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Molecular Systems Inc
- Product
- Amplilink Software versions 1.1, 1.3, 1.4, 2.41. COBAS Amplicor CTA & NGA Tests
- Product code
- JJF Analyzer, Chemistry, Micro, For Clinical Use
- Reason for recall
- For select COBAS AMPLICOR Tests run on the COBAS AMPLICOR Analyzer in conjunction with AMPLILINK software, a discrepancy has been identified between the onboard working reagent stability information reported by AMPLILINK Software (versions 1.1, 1.3, 1.4 and 2.41) and the stability information provided in the test kit package inserts/method manual.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Urgent Medical Device Correction letters were mailed on 5/21/07 by first class mail. Letters describe the discrepancy between software reagent expiration dates and package insert expiration dates.
- Quantity in commerce
- 5,806 total
- Distribution
- Nationwide to hospitals and laboratories.
- Product codes and lots
- Catalog number 20757497122, working probe suspension /IC PS1-IC4; Catalog Number 20757535122, Working substrate SB3-SB