Z-0926-2007

NeatNick Sweeping Action Heel Lancet -Full-Term- Item Number: 1030085 (box of 100) and Item Number: 1030086...

Hawaii Medical LLC / initiated 2007-05-11 / Terminated

Hawaii Medical LLC began a recall of NeatNick Sweeping Action Heel Lancet -Full-Term- Item Number: 1030085 (box of 100) and Item Number: 1030086, case of 1,000 (10 boxes of 100) on . The reason FDA records is "Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0926-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hawaii Medical LLC
Product
NeatNick Sweeping Action Heel Lancet -Full-Term- Item Number: 1030085 (box of 100) and Item Number: 1030086, case of 1,000 (10 boxes of 100)
Product code
FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature
Reason for recall
Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient
FDA root cause
Other
Action
Hawaii Medical notified the initial distributor and end users by telephone and follow-up Recall Notification Letter on May 15, 2007.
Quantity in commerce
286,376 units
Distribution
Nationwide
Product codes and lots
M073010, M085019, M088037, M094012, M106010, M107016, M116013, M122004