Z-0927-2007
NeatNick Sweeping Action Heel Lancet - Preemie- Item Number: 1030083 (box of 100) and Item Number: 1030084...
Hawaii Medical LLC / initiated 2007-05-11 / Terminated
Hawaii Medical LLC began a recall of NeatNick Sweeping Action Heel Lancet - Preemie- Item Number: 1030083 (box of 100) and Item Number: 1030084, case of 1,000 (10 boxes of 100) on . The reason FDA records is "Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0927-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hawaii Medical LLC
- Product
- NeatNick Sweeping Action Heel Lancet - Preemie- Item Number: 1030083 (box of 100) and Item Number: 1030084, case of 1,000 (10 boxes of 100)
- Product code
- FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature
- Reason for recall
- Lancet may misfire and an inadequate blood sample may result; additionally, the blade may not retract after use and result in a needle stick to the user/patient
- FDA root cause
- Other
- Action
- Hawaii Medical notified the initial distributor and end users by telephone and follow-up Recall Notification Letter on May 15, 2007.
- Quantity in commerce
- 58,058 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Number: M095007, M066008