Precision Bipolar Device ClearGlide EVH Small (endoscopic vessel harvesting system) The kit included three...
Datascope Corporation began a recall of Precision Bipolar Device ClearGlide EVH Small (endoscopic vessel harvesting system) The kit included three Precision BiPolar Devices, an Optical Vessel... on . The reason FDA records is "Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device. Reports include the locking or sticking of the buttons of the device which operate the jaws, particularly when a...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1156-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- Precision Bipolar Device ClearGlide EVH Small (endoscopic vessel harvesting system) The kit included three Precision BiPolar Devices, an Optical Vessel Dissector, a Small Ultra Retractor and a Vessel Dissector. Datascope Cardiac Assist Sterile R Manufactured for: Datascope Cardiac Assist 15 Law Drive Fairfield, NJ 07004
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Datascope has experienced a higher than usual number of reports involving the Precision Bipolar Device. Reports include the locking or sticking of the buttons of the device which operate the jaws, particularly when a side load is placed on the shaft of the device.
- FDA root cause
- Other
- Action
- Letters were sent to all hospitals, (contact: Hospital Administrators) the week of May 7, 2007 (certified mail, Return Receipt requested) with Urgent Medical Device Field Correction. The letter requests that all potential users of the devices at the institution be mdae aware of the situation. The letter instructs clinicians what to do if they experience problems with locking buttons of the Precision Device and also if the device is clamped closed on a vein branch and cannot be opened. Institutions are also requested to sign and return a postal card acknowledging receipt of the Field Correct notification and Datascope provides telephone numbers for customer service.
- Quantity in commerce
- 1008
- Distribution
- Worldwide, including USA, Belgium, the Netherlands, and Germany.
- Product codes and lots
- Part number KTV 15 Lot numbers: 12499, Exp. 4/21/09; 12500, Exp. 4/22/09; 12501, Exp. 4/27/09; 12504, Exp. 5/04/09 and 12505, Exp. 5/13/09.