OneTouch Ultra Test strip, a component of the following OneTouch Ultra Glucose monitoring systems: a)OneTouch...
Lifescan Inc began a recall of OneTouch Ultra Test strip, a component of the following OneTouch Ultra Glucose monitoring systems: a)OneTouch Ultra, b) OneTouch InDuo, c) OneTouch Ultra2, d)... on . The reason FDA records is "Damaged Vials/Inaccurate test results: a manufacturing equipment problem on one of the packaging lines had the potential to puncture vials, causing a single hole in the side of the vial below the rim. This result of...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1042-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lifescan, Inc.
- Product
- OneTouch Ultra Test strip, a component of the following OneTouch Ultra Glucose monitoring systems: a)OneTouch Ultra, b) OneTouch InDuo, c) OneTouch Ultra2, d) OneTouch UltraMini, e) OneTouch UltraSmart. 25 strips per vial. Product comes in packs of 100 and 50. (Part numbers: 020-245-07, 020-244-07, 020-963-01)
- Product code
- NBW System, Test, Blood Glucose, Over The Counter
- Reason for recall
- Damaged Vials/Inaccurate test results: a manufacturing equipment problem on one of the packaging lines had the potential to puncture vials, causing a single hole in the side of the vial below the rim. This result of this defect may produce inaccurately high or low blood glucose test results or error message readings.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- March 2007 via letters to consumers and consignees and web postings. The firm has also set up verbal communication plans for customers who contact Lifescan by phone.
- Quantity in commerce
- ~33 million vials (25 strips per vial in the US, ~19 million worldwide.
- Distribution
- Worldwide; USA, Europe, Middle East, Africa, Latin America and Canada.
- Product codes and lots
- Part numbers 020-245-07 (100 pack), 020-244-07 (50 pack), 020-963-01 (50 pack mail order). Expiration dates 08/2007 through 11/2008.