Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm)

STERIS Corporation / initiated 2007-03-22 / Terminated / root cause: manufacturing process

Steris Corp began a recall of Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm) on . The reason FDA records is "The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1193-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).
Product code
FLE Sterilizer, Steam
Reason for recall
The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
By telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.
Quantity in commerce
78
Distribution
Worldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
Product codes and lots
Prevac Steam Sterilizer, Model #'s V-136H, V-148H, V-160H, SV-148H, V-1264 & V-1263& SFPP Steam Sterilizer, Model #'s P-148H & P-160H.   Code # B093934202, B093924204, B09324223, B093924225, B093924227, B093924229, B093924233, B093924244, B093924246, B093924248, B093924250, B093924286, B093924288, B093924296, B093924307, B093924309, B093924310, B093924434, B093924476, B093927301, B093927302, B093927322, B300903936