Z-1193-2007
Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm)
STERIS Corporation / initiated 2007-03-22 / Terminated / root cause: manufacturing process
Steris Corp began a recall of Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm) on . The reason FDA records is "The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1193-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).
- Product code
- FLE Sterilizer, Steam
- Reason for recall
- The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- By telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.
- Quantity in commerce
- 78
- Distribution
- Worldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
- Product codes and lots
- Prevac Steam Sterilizer, Model #'s V-136H, V-148H, V-160H, SV-148H, V-1264 & V-1263& SFPP Steam Sterilizer, Model #'s P-148H & P-160H. Code # B093934202, B093924204, B09324223, B093924225, B093924227, B093924229, B093924233, B093924244, B093924246, B093924248, B093924250, B093924286, B093924288, B093924296, B093924307, B093924309, B093924310, B093924434, B093924476, B093927301, B093927302, B093927322, B300903936