Z-1115-2007
ThromCat Thrombectomy Catheter System. Catalog number 63000-01
DSM Biomedical, Inc. / initiated 2007-05-21 / Terminated
Kensey Nash Corp began a recall of ThromCat Thrombectomy Catheter System. Catalog number 63000-01 on . The reason FDA records is "Defective component (power supply)". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1115-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DSM Biomedical, Inc.
- Product
- ThromCat Thrombectomy Catheter System. Catalog number 63000-01
- Product code
- MCW Catheter, Peripheral, Atherectomy
- Reason for recall
- Defective component (power supply)
- FDA root cause
- Other
- Action
- New power supplies were sent directly to Sales Representatives who personally replaced the defective power supplies and returned them to Kensey Nash. A voice mail to all affected KNC Sales Representatives went out on 5/21/07 instructing them to contact the affected accounts immediately.
- Quantity in commerce
- 11 units
- Distribution
- Nationwide:The products were shipped to medical facilities in FL, MO, NJ, and TX.
- Product codes and lots
- Catalog number 63000-01, Lot number 57292 exp March 2008