Z-1069-2007

IMx Total B-hCG Reagent is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti...

Abbott Diagnostic International, Ltd. / initiated 2007-04-10 / Terminated

Abbott Diagnostic International, Ltd. began a recall of IMx Total B-hCG Reagent is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti B-hCG Alkaline Phosphatase Conjungate, one bottle of... on . The reason FDA records is "An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1069-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Diagnostic International, Ltd.
Product
IMx Total B-hCG Reagent is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti B-hCG Alkaline Phosphatase Conjungate, one bottle of 4-Methylumbelliferyl Phosphate and one bottle of Specimen Diluent.
Product code
JHI Visual, Pregnancy Hcg, Prescription Use
Reason for recall
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.
FDA root cause
Other
Action
The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots.
Quantity in commerce
3176 kits
Distribution
Worldwide, including USA, PR, USVI, CANADA, MEXICO, GUATEMALA, COSTA RICA, COLOMBIA, ECUADOR, PERU, CHILE, URUGUAY, ARGENTINA, BRAZIL, DOMINICAN REPUBLIC, EL SALVADOR, GERMANY, JAPAN, SINGAPORE, HONG KONG, TAIWAN, THAILAND, SOUTH KOREA, AUSTRALIA, NEW ZEALAND, HONDURAS, TRINIDAD & TOBAGO, GRENADA, ST. VINCENT, ST. LUCIA, ANTIGUA, BARBADOS, CURACAO, BRITISH VIRGIN ISLANDS, BAHAMAS, PANAMA, and CAYMAN ISLANDS.
Product codes and lots
46810Q100, 49118Q100, 49555Q100