Z-1185-2007
AxSYM Quinidine Standard Calibrators - consist of six bottles of Cal A (6ml) and Cals B-F (4ml each)-Product...
Abbott Diagnostic International, Ltd. / initiated 2007-05-31 / Terminated
Abbott Diagnostic International, Ltd. began a recall of AxSYM Quinidine Standard Calibrators - consist of six bottles of Cal A (6ml) and Cals B-F (4ml each)-Product List Number-7A73-01 on . The reason FDA records is "Controls Out of Range/Specification". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1185-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Diagnostic International, Ltd.
- Product
- AxSYM Quinidine Standard Calibrators - consist of six bottles of Cal A (6ml) and Cals B-F (4ml each)-Product List Number-7A73-01
- Product code
- LGI Fluorescence Polarization Immunoassay, Carbamazepine
- Reason for recall
- Controls Out of Range/Specification .
- FDA root cause
- Other
- Action
- May 31, 2007, Product Correction Letter to Customers were instructed to: "Revise the expiration date for all lots identified in the customer letter. "Discontinue use of the expired lots, and discard the material according to their laboratory procedures. "An additional option is to use the Abbott Immunoassay Multiconstituent Control (List Number 06E20), manufactured by BIO-RAD.
- Quantity in commerce
- 267
- Distribution
- Worldwide: USA including states of AR, AZ, CA, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, NV, OH, OK, OR, SC, TN, TX, VA, WA, WI, WV, Hawaii and Puerto Rico. International: CANADA, GERMANY, JAPAN, SINGAPORE, AUSTRALIA, and BARBADOS
- Product codes and lots
- Lot # 44794Q100 Exp. Date: 08/09/2007 and Lot # 47093Q100 Exp. Date: 10/10/2007