Z-1186-2007

AxSYM Quinidine Controls - are composed of 3 bottles of controls (AxSYM 8mL & XSYSTEM 2.5mL -Control Low...

Abbott Diagnostic International, Ltd. / initiated 2007-05-31 / Terminated

Abbott Diagnostic International, Ltd. began a recall of AxSYM Quinidine Controls - are composed of 3 bottles of controls (AxSYM 8mL & XSYSTEM 2.5mL -Control Low, Control Medium, Control High) containing quinidine... on . The reason FDA records is "Controls Out of Range/Specification". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1186-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Diagnostic International, Ltd.
Product
AxSYM Quinidine Controls - are composed of 3 bottles of controls (AxSYM 8mL & XSYSTEM 2.5mL -Control Low, Control Medium, Control High) containing quinidine prepared in human serum to yield the different concentration ranges-Product List Number 7A73-10
Product code
LGI Fluorescence Polarization Immunoassay, Carbamazepine
Reason for recall
Controls Out of Range/Specification .
FDA root cause
Other
Action
May 31, 2007, Product Correction Letter to Customers were instructed to: "Revise the expiration date for all lots identified in the customer letter. "Discontinue use of the expired lots, and discard the material according to their laboratory procedures. "An additional option is to use the Abbott Immunoassay Multiconstituent Control (List Number 06E20), manufactured by BIO-RAD.
Quantity in commerce
119
Distribution
Worldwide: USA including states of AR, AZ, CA, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, NV, OH, OK, OR, SC, TN, TX, VA, WA, WI, WV, Hawaii and Puerto Rico. International: CANADA, GERMANY, JAPAN, SINGAPORE, AUSTRALIA, and BARBADOS
Product codes and lots
Lot # 43273Q100 Exp. Date: 06/23/2007 and  Lot # 47028Q100 Exp. Date; 10/14/2007