Z-1077-2007
DCI Equipment Track Mount Light Assembly, 110v
DCI International, LLC / initiated 2007-05-01 / Terminated
Dental Equipment LLC began a recall of DCI Equipment Track Mount Light Assembly, 110v on . The reason FDA records is "Potential for light shield to melt due to light assemblies not properly rated to handle the VAC". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1077-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- DCI International, LLC
- Product
- DCI Equipment Track Mount Light Assembly, 110v
- Product code
- EAZ Light, Operating, Dental
- Reason for recall
- Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.
- FDA root cause
- Other
- Action
- On 5/1/07 the dental office contacted the manufacturer via telephone regarding the incident. On 5/2/07 the firm's service representative visited the dental office and replaced the lights.
- Quantity in commerce
- 4 dual track lights
- Distribution
- The lights were distributed to one distributor in NY who in turn distributed the lights to one dental office.
- Product codes and lots
- All products of this type. Product does not contain serial numbers or any other identifying codes.