Z-1102-2007
Cell Search Circulating Tumor Cell Kit (Epithelial) Veridex LLC, a Johnson & Johnson Company IVD Veridex, LLC...
Veridex, LLC / initiated 2007-04-17 / Terminated
Veridex, LLC began a recall of Cell Search Circulating Tumor Cell Kit (Epithelial) Veridex LLC, a Johnson & Johnson Company IVD Veridex, LLC 33 Technology Drive Warren, NJ 07059 USA... on . The reason FDA records is "An increased incidence of aggregates was observed in Lot 0036 and 0036A. The presence of aggregates will cause customers to invalidate patient results". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1102-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Veridex, LLC
- Product
- Cell Search Circulating Tumor Cell Kit (Epithelial) Veridex LLC, a Johnson & Johnson Company IVD Veridex, LLC 33 Technology Drive Warren, NJ 07059 USA 1-877-VERIDEX USA
- Product code
- NQI System, Immunomagnetic, Circulating Cancer Cell, Enumeration
- Reason for recall
- An increased incidence of aggregates was observed in Lot 0036 and 0036A. The presence of aggregates will cause customers to invalidate patient results.
- FDA root cause
- Other
- Action
- All Direct customers and customers of distributors received a customer notification informing them of the increased incidence of aggregates in cartridges on April 30, 2007 by Fed Ex overnight. This letter instructed them to discontinue use and to discard any remaining inventory. This was followed by a notification on May 24, 2007 by Fed Ex overnight to all US Distributors which also requested that they immediately stop using and/or distributing kits and to destroy any remaining inventory. Distributors were also asked to notifiy customers who received product. All foreign affiliate consignees were also informed on this date by e-mail and instructed to notify their customers.
- Quantity in commerce
- Lot 0036: 468 kits; Lot 0036A: 197 kits
- Distribution
- The kits were distributed to hospitals and laboratories nationwide. They were also sent to 7 J&J affiliated companies in France, Japan, Canada, England, Germany, Italy and Spain.
- Product codes and lots
- Ref 7900001, Lot 0036 Ref 7900000, Lot 0036A - This is a research use only lot and not for in vitro diagnostic use.