VITROS 5,1 FS (Fusion Series) Chemistry System Software Version 2.2.1, only when using VITROS Chemistry...
Ortho-Clinical Diagnostics began a recall of VITROS 5,1 FS (Fusion Series) Chemistry System Software Version 2.2.1, only when using VITROS Chemistry Products THC (Cannabinoids) or AMPH (Amphetamine)... on . The reason FDA records is "Software design. The firm received a customer complaint regarding a qualitative positive THC (Cannabinoids) result obtained from a THC quality control fluid that was known to be negative when using VITROS Chemistry...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1167-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- VITROS 5,1 FS (Fusion Series) Chemistry System Software Version 2.2.1, only when using VITROS Chemistry Products THC (Cannabinoids) or AMPH (Amphetamine) Reagents, Cat. No. 680 1375.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Software design. The firm received a customer complaint regarding a qualitative positive THC (Cannabinoids) result obtained from a THC quality control fluid that was known to be negative when using VITROS Chemistry Products THC Reagent. A second customer complaint was received a month later. OCD's investigation concluded that under specific circumstances, incorrect qualitative results for Cannabi
- Root cause tag
- software
- FDA root cause
- Other
- Action
- On 5/17/07, OCD informed its foreign affiliates of the field correction by e-mail, and instructed them to inform their customers who received the VITROS 5,1 FS Chemistry Systems as well as the VITROS Chemistry Products THC (Cannabinoids) and/or AMPH (Amphetamine) Reagents of this issue. On 5/18/07, all US OCD direct customers and end-use customers of consignees that received the VITROS 5,1 FS Chemistry Systems as well as the VITROS Chemistry Products THC (Cannabinoids) and/or AMPH (Amphetamine) Reagents, were informed of the "field correction" by US Postal Service letter, provided with a new Assay Data Disk (ADD #5465), and were instructed to load it and recalibrate these assays, if necessary.
- Quantity in commerce
- Domestic: 604, International: 598
- Distribution
- Nationwide and Worldwide
- Product codes and lots
- Software Version 2.2.1, Serial Numbers 34000121 to 34001377.