Z-1018-2007
Nonabsorbable Polypropylene Synthetic Suture-Deklene, DEK BL MF 0 TC-43 2N 48, Catalog #833-123
Teleflex Medical / initiated 2007-06-06 / Terminated
Teleflex Creek Dr began a recall of Nonabsorbable Polypropylene Synthetic Suture-Deklene, DEK BL MF 0 TC-43 2N 48, Catalog #833-123 on . The reason FDA records is "Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1018-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Nonabsorbable Polypropylene Synthetic Suture-Deklene, DEK BL MF 0 TC-43 2N 48, Catalog #833-123
- Product code
- GAM Suture, Absorbable, Synthetic, Polyglycolic Acid
- Reason for recall
- Suture Diameter/Strength specification failure: The products may not meet their final specifications for suture diameter and/or strength which could lead to premature breakage.
- FDA root cause
- Other
- Action
- Consignees were notified by letter on/about 06/07/2007. They were advised to cease use, remove and quarantine all affected product immediately. Affected products are to be returned to Teleflex Medical. An Acknowledgement & Stock form was included to be completed and returned via fax to Teleflex Medical.
- Quantity in commerce
- 1920 units total for all types of sutures
- Distribution
- Worldwide-, USA including states of CO, FL, IL, IN, MA, and OH, and Countries of Canada, France and Germany
- Product codes and lots
- Deklene, DEK BL MF 0 TC-43 2N 48, Catalog #833-123, Lot #35089.