Z-1142-2007
The Companion ClearStar Pump System, for enteral feeding
Ross Products Division Abbott Laboratories / initiated 2007-06-21 / Terminated / root cause: manufacturing process
Ross Products Division Abbott Laboratories began a recall of The Companion ClearStar Pump System, for enteral feeding on . The reason FDA records is "The transducer adapter bracket on the pump could crack or break and potentially result in under delivery of the enteral product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1142-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ross Products Division Abbott Laboratories
- Product
- The Companion ClearStar Pump System, for enteral feeding.
- Product code
- LZH Pump, Infusion, Enteral
- Reason for recall
- The transducer adapter bracket on the pump could crack or break and potentially result in under delivery of the enteral product.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- Consignees were notified by letter dated 6/25/2007. A revised letter was sent 07/26/2007.
- Quantity in commerce
- 27,814
- Distribution
- Worldwide, including USA and Canada.
- Product codes and lots
- List #55238 & 55239