Z-0680-2007
OEC FluoroTrak 9800 PLUS, image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT
GE Healthcare / initiated 2007-03-01 / Terminated
GE OEC Medical Systems, Inc began a recall of OEC FluoroTrak 9800 PLUS, image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT on . The reason FDA records is "During routine service, primary and/or secondary collimators may not have been properly reinstalled". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0680-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- OEC FluoroTrak 9800 PLUS, image-intensified fluoroscopic x-ray system, GE Healthcare, Salt Lake City, UT.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- During routine service, primary and/or secondary collimators may not have been properly reinstalled.
- FDA root cause
- Other
- Action
- Customers were notified by letter on 03/01/2007. They were told to continue using the machines and that an inspection would be scheduled immediately to verify that collimators had been correctly installed.
- Quantity in commerce
- 5 units
- Distribution
- Nationwide and Germany.
- Product codes and lots
- Serial numbers: 89-1579-N, 89-2529-N, 82-2160-N, 8S-1932-N, 8S-2226-N.