Z-0717-2007
Kodak Direct View DR3000 System, Stationary X-ray System, Catalog #s 655-1345, 655-1386
Eastman Kodak Company / initiated 2007-03-06 / Terminated
Carestream Health, Inc. began a recall of Kodak Direct View DR3000 System, Stationary X-ray System, Catalog #s 655-1345, 655-1386 on . The reason FDA records is "The x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0717-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Eastman Kodak Company
- Product
- Kodak Direct View DR3000 System, Stationary X-ray System, Catalog #s 655-1345, 655-1386.
- Product code
- KPR System, X-Ray, Stationary
- Reason for recall
- The x-ray systems do not indicate the field size when the system's detector is turned off and the collimator is in manual mode. This results in the systems being noncompliant with the field indication performance standards of section 1020.31(e)(2)& (3) of 21Code of Federal Regulations (CFR).
- FDA root cause
- Other
- Action
- On 3/06/07, Eastman Kodak notified CDRH of its proposed Corrective Action Plan (CAP). The CAP was approved by CDRH under the following conditions: (1) Kodak will visit each user facility, and apply dial indicator labels to the collimator knobs, apply a field size label to the collimator housing, and verify that each system complies with the field indication performance standard. (2) Kodak will prepare adequate instructions for the assembly/installation and testing of the dial indicators labels and field size labels. These instructions are to be furnished to all owners of the Kodak Direct View DR3000 Systems. (3) Kodak will supplement the Kodak DR3000 system product report with a copy of the adequate instructions. (4) Kodak will perform 100% verification of the recall effectiveness. (5) Kodak will not charge for the corrections and will reimburse the owner for any out-of-pocket expenses. (6) Kodak will notify the manufacturer of the Ralco 302 ACS collimators, Ralco, that the collimator is noncompliant with the performance standard.
- Quantity in commerce
- Domestic: 7 units, Canada: 2 units
- Distribution
- FL, NY, MO, MI, TX, and Canada
- Product codes and lots
- Serial Numbers DR30001001, DR30001004, DR30001006, DR30001014, DR30001016, DR30001017, DR30001020, DR30001021, DR30000032.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
