Z-1086-2007
SPI EASY Abutment. Endosseous dental implant abutment
Thommen Medical AG / initiated 2007-04-19 / Terminated
Thommen Medical, AG began a recall of SPI EASY Abutment. Endosseous dental implant abutment on . The reason FDA records is "The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1086-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Thommen Medical AG
- Product
- SPI EASY Abutment. Endosseous dental implant abutment.
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality expectations of the firm.
- FDA root cause
- Other
- Action
- Letter 5/14/2007. Customers are requested to check their current inventory for the recall product. If found they are to return the product for replacement. If they have implanted the product they are to check patient x-rays for the defect. If defects are found they are to do a restoration.
- Quantity in commerce
- 84
- Distribution
- Nationwide.
- Product codes and lots
- #4_03_700 Batch 8007 REF: 1.03.700 LOT: 5121
FIRST ARTICLE / WEEKLY