Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1086-2007

SPI EASY Abutment. Endosseous dental implant abutment

Thommen Medical AG / initiated 2007-04-19 / Terminated

Thommen Medical, AG began a recall of SPI EASY Abutment. Endosseous dental implant abutment on . The reason FDA records is "The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1086-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Thommen Medical AG
Product
SPI EASY Abutment. Endosseous dental implant abutment.
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality expectations of the firm.
FDA root cause
Other
Action
Letter 5/14/2007. Customers are requested to check their current inventory for the recall product. If found they are to return the product for replacement. If they have implanted the product they are to check patient x-rays for the defect. If defects are found they are to do a restoration.
Quantity in commerce
84
Distribution
Nationwide.
Product codes and lots
#4_03_700 Batch 8007 REF: 1.03.700 LOT: 5121