Z-0848-2007
GE OEC 9900 Elite Fluoroscopic X-ray System, GE Healthcare, Salt Lake City, Utah
GE Healthcare / initiated 2007-06-12 / Terminated
GE OEC Medical Systems, Inc began a recall of GE OEC 9900 Elite Fluoroscopic X-ray System, GE Healthcare, Salt Lake City, Utah on . The reason FDA records is "X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0848-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE OEC 9900 Elite Fluoroscopic X-ray System, GE Healthcare, Salt Lake City, Utah.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- X-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma.
- FDA root cause
- Other
- Action
- Consignees were notified by letter on 06/12/2007 and told that a service representative would arrange for an inspection and would modify their systems.
- Quantity in commerce
- 295 units
- Distribution
- Worldwide Distribution: USA including states of FL, GA, PA. CO, MD, VA and WA. and countries of Australia, Austria, Belgium, Brazil, Canada, Colombia, Finland, France, Germany, Hong Kong, India, Israel, Italy, Japan, Korea, Netherlands, New Zealand, Norway, Portugal, Switzerland, Taiwan, and the UK.
- Product codes and lots
- Serial numbers: E2-0155, E2-0156, E2-0157, E2-0164, E2-0167, E2-0168, E2-0170 thru E2-0234, E2-0235-C, E2-0236 thru E2-0257, E2-0259, E2-7048-MH, E2-7050-MH, E2-7051-MH, E2-7058-MH, E2-7059-MH, E2-7060-MH, E2-7061-MH, E2-7062-MH, E2-7064-MH, E2-7065-MH, E2-7066-MH, E2-7067-MH, E2-7068-MH, E2-7069-CMH, E2-7070-MH, E2-7071-MH, E2-7072-MH, E2-7073-MH, E2-7074-MH, E2-7076-MH, E2-7077-CMH, E9-0037, E9-0085, E9-0088, E9-0089-G, E9-0090, E9-0093, E9-0094, E9-0095, E9-0098 thru E9-0103, E9-0108 thru E9-0153, E9-0155, E9-0156, E9-0157, E9-0158, E9-0159, E9-0162, E9-0163, E9-0164, E9-0165, E9-0170, E9-0171, E9-0172, E9-0173, E9-0178, E9-0183, ES-0041-G, ES-0069, ES-0071, ES-0072, ES-0078, ES-0084-G, ES-0086-G, ES-0087, ES-0088, ES-0089, ES-0090, ES-0092-CH, ES-0096, ES-0097, ES-0098, ES-0099-CH, ES-0101-CH, ES-0103, ES-0104, ES-0106 thru ES-0130, ES-0131-H, ES-0132 thru ES-0154, ES-0155-H, ES-0156, ES-0157, ES-0158, ES-0159, ES-0160, ES-0162-G thru ES-0167-G, ES-0169-CH, ES-0170, ES-0171, ES-0172, ES-0173, ES-0175, ES-0176, ES-0177, ES-0178, ES-0184, ES-7018-MH, ES-7026-CMH thru ES-7031-CMH, ES-7032-MH, ES-7033-CMH, ES-7034-CMH, ES-7035-CMH, ES-7036-CMH.