Z-1245-2007
daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly)...
Intuitive Surgical, Inc. / initiated 2007-05-14 / Terminated / root cause: manufacturing process
Intuitive Surgical, Inc. began a recall of daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system on . The reason FDA records is "Product may malfunction and fail to start up on AC power". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1245-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- daVinci S Surgical system, patient side cart Auxiliary power board (APB) (printed circuit board assembly), endoscopic instrument control system.
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Product may malfunction and fail to start up on AC power.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- As of 9/12/2007, the firm has conducted a field inspection of any involved units via service visit. No other notice is planned or indicated by the firm's submission. The firm feels that the risk associated with this issue is minimal, and therefore, any advanced notification is unneccessary. If any system failed one or more of the inspection points, its APB (3510404-04) was replaced with a verified APB (351040-04 or 351040-05).
- Quantity in commerce
- 38 units
- Distribution
- Worldwide, including USA, France, China, Japan, Mexico, Spain, Italy, Belgium and Singapore.
- Product codes and lots
- Model number IS2000, part number 351040-04. System numbers SG002, SG004, SG010 , SG032, SG060, SG061, SG064, SG065, SG066, SG067, SG068, SG069, SG070, SG071, SG072, SG073, SG074, SG075, SG076, SG077, SG078, SG079, SG080, SG081, SG082, SG083, SG084, SG086, SG090, SG116, SG117, SG118, SG120, SG121, SG126, SG174, SG175, SG184