Z-0242-2008
OSI ProFx Operating Room Table 6850 , Pelvic Reconstruction Orthopedic Surgery Table-- Stand alone...
Mizuho Orthopedic Systems Inc / initiated 2007-02-28 / Terminated / root cause: software
Mizuho Orthopedic Systems Inc began a recall of OSI ProFx Operating Room Table 6850 , Pelvic Reconstruction Orthopedic Surgery Table-- Stand alone single-pedestal AC power or battery operated table... on . The reason FDA records is "Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0242-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mizuho Orthopedic Systems Inc
- Product
- OSI ProFx Operating Room Table 6850 , Pelvic Reconstruction Orthopedic Surgery Table-- Stand alone single-pedestal AC power or battery operated table, Orthopedic Systems, Inc., Union City, CA
- Product code
- JEA Table, Surgical With Orthopedic Accessories, Ac-Powered
- Reason for recall
- Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.
- Quantity in commerce
- 109 units
- Distribution
- Worldwide: USA product was distributed to 16 wholesale dealers, 212 medical facilities, 14 distributors throughout the US in the following states: AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, WI, WV and WY and to the following countries Taiwan, Australia, German, Japan, Canada, Saudi Arabia, and The UK.
- Product codes and lots
- Serial Numbers: 1-12, 14-21, 23,24, 26-39, 41-43, 45-54, 56-59, 62-76, 78-100, and 102-119