Z-0243-2008
OSI, 6822 IntraOp Series Tables, Operating Room tables, Orthopedic Systems, Inc., Union City, CA. (Series...
Mizuho Orthopedic Systems Inc / initiated 2007-02-28 / Terminated / root cause: software
Mizuho Orthopedic Systems Inc began a recall of OSI, 6822 IntraOp Series Tables, Operating Room tables, Orthopedic Systems, Inc., Union City, CA. (Series 6800 and 6822 included) on . The reason FDA records is "Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0243-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Mizuho Orthopedic Systems Inc
- Product
- OSI, 6822 IntraOp Series Tables, Operating Room tables, Orthopedic Systems, Inc., Union City, CA. (Series 6800 and 6822 included)
- Product code
- LHN System, Radiation Therapy, Charged-Particle, Medical
- Reason for recall
- Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.
- Quantity in commerce
- 117 units
- Distribution
- Worldwide: USA product was distributed to 16 wholesale dealers, 212 medical facilities, 14 distributors throughout the US in the following states: AL, AZ, CA, FL, GA, HI, IL, IN, KS, KY, MA, MD, ME, MI, MO, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, VA, WA, WI, WV and WY and to the following countries Taiwan, Australia, German, Japan, Canada, Saudi Arabia, and The UK.
- Product codes and lots
- Serial Numbers: 12-15, 17, 19, 21-57, 59-61, 64-67, 70, 71, 73-83, 85-93, 95, 97, 99-134 and 136-141