Z-1190-2007
Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control...
Volcano Corporation / initiated 2007-04-02 / Terminated
Volcano Corporation began a recall of Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module... on . The reason FDA records is "During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1190-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Volcano Corporation
- Product
- Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)
- Product code
- IYO System, Imaging, Pulsed Echo, Ultrasonic
- Reason for recall
- During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.
- FDA root cause
- Other
- Action
- The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.
- Quantity in commerce
- 351 units worldwide
- Distribution
- Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).
- Product codes and lots
- Device listing A863634, part number 804200-100. monitor part number that is being replaced: 90X0315A. All units shipped on or before November 2005.