Z-1190-2007

Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control...

Volcano Corporation / initiated 2007-04-02 / Terminated

Volcano Corporation began a recall of Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module... on . The reason FDA records is "During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1190-2007
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Volcano Corporation
Product
Volcano s5 Imaging system, Ultrasound pulsed echo imaging system. Product consists of the CPU, control console, monitor, printer, and patient interface module (PIM)
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
During emissions testing, it was discovered that the product exceeds emissions limits for electromagnetic compatibility.
FDA root cause
Other
Action
The firm notified its customers by letter beginning April 2, 2007, and started replacing the monitor with appropriate replacement parts.
Quantity in commerce
351 units worldwide
Distribution
Product was distributed to health care facilities nationwide and to distributors in Europe (via subsidiary), Japan (via subsidiary), Latin America and Asia (third party distributors).
Product codes and lots
Device listing A863634, part number 804200-100. monitor part number that is being replaced: 90X0315A. All units shipped on or before November 2005.