Z-0130-2008
GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075...
American Medical Systems Innovation Center - Silicon Valley / initiated 2007-04-17 / Terminated / root cause: manufacturing process
AMS Innovative Center - San Jose began a recall of GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075 (international units), AMS Innovative Center, San Jose, CA... on . The reason FDA records is "Component missing: Some products were issued without thermal protection switches". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0130-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- American Medical Systems Innovation Center - Silicon Valley
- Product
- GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075 (international units), AMS Innovative Center, San Jose, CA 95134.
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Component missing: Some products were issued without thermal protection switches.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- AMS issued Important: Medical Device Correction letters, dated April 17, 2007, to its US and international consignees by mail, instructing them to return an acknowledgement form. The letter informs customers that some HPS systems may lack a motor overload switch, that the firm believes that there is no immediate safety issue, and installation of the MOL switch will be conducted during normal preventive maintenance. AMS expects to complete installation of the switches by end of August 2007.
- Quantity in commerce
- 169 units
- Distribution
- Worldwide Distribution - USA, Italy, France, Germany, Spain, UK, Denmark, Sweden, Australia, India, Japan, Greece, Switzerland, Saudi Arabia, Thailand, Turkey, Singapore, Taiwan/Hong Kong (China), and Austria.
- Product codes and lots
- Part number 001-0070. International parts get a part number designation of 0010-0075. Serial Numbers (All starting with HPS2): 125, 167, 057, 067, 072, 106, 114, 120, 149, 193, 186, 161, 037, 039, 040, 044, 050, 088, 185, 190, 197, 035, 160, 208, 046, 076, 102, 074, 108, 121, 141, 173, 178, 089, 069, 030, 155, 172, 174, 148, 042, 164, 098, 175, 100, 169, 196, 056, 071, 132, 165, 036, 201, 033, 123, 065, 206, 170, 179, 095, 177, 073, 194, 086, 064, 122, 133, 049, 060, 078, 112, 150, 163, 130, 059, 101, 113, 135, 146, 137, 189, 195, 063, 157, 181, 166, 052, 043, 061, 092, 124, 029, 096, 093, 094, 138, 202, 211, 129, 144, 058, 162, 168, 045, 085, 066, 082, 053, 140, 158, 159, 090, 182, 183, 051, 087, 104, 116, 117, 205, 034, 203, 222, 223, 226, 229, 207, 214, 220, 230, 070, 103, 215, 110, 068, 077, 079, 128, 153, 075, 080, 083, 084, 097, 176, 105, 134, 136, 143, 147, 111, 171, 131, 151, 156, 109, 180, 145, 062, 118, 126, 127, 152, 139, 142, 209, 081, 099, 119.
FIRST ARTICLE / WEEKLY