FACSDiva Software Versions 4.0.1 to 5.0.2, Flow Cytometry Acquisition & Analysis software. Catalog number...
BD Biosciences began a recall of FACSDiva Software Versions 4.0.1 to 5.0.2, Flow Cytometry Acquisition & Analysis software. Catalog number 641609 Flexible data acquistion and analysis package... on . The reason FDA records is "When data file containing one or no fluoresence parameters is exported, the software will automatically apply compensation to this file and all subsequently exported files". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1525-2008
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Becton, Dickinson and Company, BD Biosciences
- Product
- FACSDiva Software Versions 4.0.1 to 5.0.2, Flow Cytometry Acquisition & Analysis software. Catalog number 641609 Flexible data acquistion and analysis package specifically designed for digital-based flow cytometers.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- When data file containing one or no fluoresence parameters is exported, the software will automatically apply compensation to this file and all subsequently exported files.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm has issued a customer Important Product Information letter via mail on 04/20/2007 to inform them of the software issues. The letter states that to avoid this problem, make sure each tube or well has at least two fluorescence parameters before exporting FCS 3.0 data files. Manually exporting tubes or wells one at a time or restarting the software after the export of an FCS data file containing one or no fluorescence parameters will prevent the problem from occurring. Customers are requested to fill out and return the response form and to contact BD Customer Support Center if there are any additional questions. A software fix is being developed in version 6.0 and a free maintenance release with no product enhancments (version 5.0.3).
- Quantity in commerce
- 1074 units
- Distribution
- Worldwide Distribution --- USA including states of NY, NM, MI, GA, MD, MA, AL, CA, VA, NC, OH, TX, CO, NJ, PA, MI, CT, IA, MO, NV, KY, IL, TN, FL, MN, FL, UT, WA, OK, CT, WI, SD, AZ, and IN, and countries of Buenos Aires, BE, Brazil, Canada, Chile, China, Hong Kong, Japan, Singapore, Taiwan, Venezuela, Mexico, Trinidad and Tobago, and Columbia.
- Product codes and lots
- Software versions 4.0.1 to 5.0.2.