ABL8O FLEX Software Version 1.10 included in the ABL8O Flex Analyzer, Model Number 914863
Sendx Medical Inc began a recall of ABL8O FLEX Software Version 1.10 included in the ABL8O Flex Analyzer, Model Number 914863 on . The reason FDA records is "The ABL8O FLEX software version 1.10 contains errors in the derived parameter calculation for cBase(B,ox) and cBase(Ecf,ox). These incorrect calculations result in a mild to moderate lowering of the base excess value...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1201-2007
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sendx Medical, Inc.
- Product
- ABL8O FLEX Software Version 1.10 included in the ABL8O Flex Analyzer, Model Number 914863
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- The ABL8O FLEX software version 1.10 contains errors in the derived parameter calculation for cBase(B,ox) and cBase(Ecf,ox). These incorrect calculations result in a mild to moderate lowering of the base excess value, the severity depending on the concentration of hemoglobin and the oxygen saturation in the sample. Elevated hemoglobin concentrations have the greatest impact on this calculation err
- Root cause tag
- software
- FDA root cause
- Other
- Action
- A total of eight (8) distributors and the Radiometer main office in Denmark were sent the recall package (Field Action Notice) by email on 6-6-07. The distributors were provided background information and specific instruction on the steps to perform to correct the problem. The Recall package included a Management Cover for Field Action Notes and the Field Action Notice. Radiometer has released a subsequent software upgrade (v.1.11) to correct the two problems noted in the Recall Notice.
- Quantity in commerce
- 165 software upgrade CDs
- Distribution
- In US to Cleveland, OH and Worldwide to Belgium, Denmark, France, Germany, The Netherlands, Sweden & United Kingdom
- Product codes and lots
- Software Version 1.10 Lot numbers: 17293, 17384, 17440, 17515, 17572 and 17663.